中文版中文版MessageContact usSitemapWelcome toShanghai MX-CEKONG Technology Co.,Ltd website!
Hotline 400 662 6286

MX-CEKONG

Your position: HOME > Product > Water Meter >
Tel: 400 662 6286
  • PH / ORP Intelligent Controller  MX-PHYB-01
PH / ORP Intelligent Controller MX-PHYB-01
MX-PHYB-01 PH/ORP tester is one of the intelligent on-line chemical analysis instruments, which is widely used in the continuous monitoring of PH or ORP value and temperature in thermal power, chemical fertilizer, metallurgy, environmental protection, pharmaceutical, biochemical, food and tap water solutions.


Tel:

400 662 6286

1.Product description:

MX-PHYB-01 PH/ORP tester is one of the intelligent on-line chemical analysis instruments, which is widely used in the continuous monitoring of PH or ORP value and temperature in thermal power, chemical fertilizer, metallurgy, environmental protection, pharmaceutical, biochemical, food and tap water solutions.
_PL_0763 0

2.Applying field:

1. Industrial automation applications
2. Wastewater treatment, waste water detection
3. Mining metallurgy, paper making, paper pulp concentration
4. Desulfurization and denitration
5. Healthcare applications, food and beverage processing industry
6. Strong acid, strong alkali, chemical industry wastewater
7. Petroleum chemical industry, textile wire drawing,etc.

3.Product feathers:

1. Card by modular design, assembly configuration is more convenient.Use a 2.4-inch 12864 dot matrix screen
2.PH/ORP measurement,temperature measurement,the upper and lower control, analog output, RS485 communication
3. Can be configured manually, automatic temperature compensation function
4. Can set high and low alarm function, and the amount of hysteresis
5. Can set the buzzer, LCD backlight switch function
 

4.Technique specifications:

Measuring range: PH(0-14 pH);ORP(-1000-+1000 mV or(-2000-+2000 mV))
Comprehensive precision: ±0.02PH;±1mV
Resolution values: 0.01 pH;1mV
Stable values: ≤0.02 pH/24H;;≤3 mV/24H
Input impedance: ≥10¹²Ω
Tem-measurement range: 0-100℃。Precision: ±0.5℃、NTC 10K
Tem-compensation: 0-100℃ Manual/automatic(NTC 10K)
Power supply: AC220V±10%(E1),50Hz; DC 24V(E2);
Relay alarm: (3A/250V/AV),
485 Communication: Compatible standard MODBUS-RTU Communication Protocol                               
Analog output: 4-20 mA output,the largest loop:750Ω, 0.1%FS

5.Ordering guidance:

Type Measuring mode Power supply Tem compensation Communication Transmitter Other requirements
MX-PH/ORPYB-01 ORP 24VDC No need to select defaulted defaulted ---
E g: MX-ORPYB-01-E2

6.Installations:

1. Cut out a 91*91 mounting hole on the instrument cabinet or installation panel
2. Insert the meter into the fixing screw in the mounting hole

7.Contact us:

Shanghai MX-CEKONG Technology Co.,Ltd

Email: 442365872@QQ.com
 
Tel: 021-67881626
 
Web: www.shmxck.com

Address: No. 488 Guanghua Road, Songjiang District, Shanghai